Environics Logo

Product Data
Series 2000
Computerized Multi-Component Gas Mixer


DESCRIPTION

The Environics® Series 2000 Computerized Multi-Component Gas Mixer automatically blends and dilutes gases to generate precise gas calibration standards, create gaseous atmospheres or produce gas mixes for analytical research or production purposes. The Series 2000 dynamically produces complex mixes containing up to eight(8) individual component gases in a balance gas. Each component concentration may be independently varied in response to user commands. The Series 2000 can produce gas concentrations from percent levels to ppb for single or multi-point calibration of gas chromagraphs, process gas analyzers, mass spectrometers, FTIR's and other gas analytic devices.

The Series 2000 consists of a single chassis supporting up to nine (9) mass flow controllers, a serpentine pre-mix zone and a zero dead-space final mixing zone. Gas wetted surfaces are electro- polished 316 stainless steel. Seals are gas-compatible elastomer. User interface includes a backlit 80 character by 25 line liquid-crystal display and membrane keypad.

The instrument's mass flow controllers are factory calibrated using a primary flow standard traceable to the United States National Institute of Standards and Technology (NIST). The calibration data consists of an eleven-point comparison of commanded versus actual flow with linear interpolation between the points.

The Series 2000 is available in either a standard rack mount or bench top configuration. The optional RS-232 serial data interface permits remote or computer operation of the instrument.


PRODUCT FEATURES AND BENEFITS



SOFTWARE:

The Series 2000 has six primary routines accessible through on screen "soft" keys.

SPECIFICATIONS

Performance (as a percent of setpoint)

Accuracy              From 10 to 100% of Full Scale Flow
   Concentration:         ± 1.0%
   Flow:                  ± 1.0%
Repeatability:             0.05%
*Mass flow controllers are calibrated using a NIST Traceable Primary Flow Standard,
using a reference temperature of 0°C (32°F) and a reference pressure of 760 mm Hg (29.92 in. Hg).

Warm up time: 30 minutes

Mechanical

Gas connections
  1/4" Swagelok (or compatible fitting)

Operating Pressures
  Minimum:       10 psig (0.67 Bar)
  Recommended:   25 psig (1.68 Bar)
  Maximum:      150 psig (9.87 Bar)
                 75 psig (5.04 Bar) with Purge option

Wetted Surfaces
  Tubing:       Electropolished 316 Stainless Steel
                (Optional - Teflon, Hastelloy, Monel)
  MFC's:        Stainless Steel (Optional - Hastelloy, Monel)
  Seals:        Viton® (Optional - Kalrez®, Buna-N, Neoprene, Metal)

Operating temperature:    32° - 122° F   (0° - 50° C)
Performance temperature:  59°  - 95° F  (15° - 35° C)

Weight
  Minimum:      35 lbs. (16 Kg.)
  Maximum:      70 lbs. (32 Kg.)

Dimensions (w x h x d)
  Portable:  17" x 7" x 23.5"
             (43.18 cm x 17.78 cm x 59.69 cm)
  Rack:      19" x 7" x 23.5"
             (48.26 cm x 17.78 cm x 59.69 cm)

Electrical

  Standard:  115 VAC (100 to 130 VAC), 50/60 Hz
  Optional:  220 VAC (200 to 260 VAC), 50/60 Hz
  Current:     3 Amps (maximum)

Electronics

  Inmos T 400 series, 32 bit processor
  12 bit A/D and D/A conversion

Operating Modes

  Front panel membrane keypad
  Internal timer control
  RS-232 terminal mode / Remote computer control (optional)
  Status board interface (optional)
  Other communications options available

Data Output

  Optional RS-232 serial port / Parallel printer port interface


OPTIONS

Viton® and Kalrez® are registered trademarks of Dupont Dow Elastomers