Computerized VOC Gas Dilution System


DESCRIPTION

The Environics® Series 2014 Computerized VOC Gas Dilution System is a microprocessor based gas dilution system designed for the preparation of calibration standards used for VOC (Volatile Organic Compounds) monitoring programs. The Series 2014 can be used to dynamically produce ppb level standards from ppm level cylinder gases.

EPA methods for VOC's specify a dynamic dilution to produce the initial multi-point calibration as well as the required daily and pre-sample calibration at the nominal 10 ppb level.

The Series 2014 employs mass flow controllers calibrated to a NIST (National Institute of Standards and Technology) traceable primary standard. All gas wetted surfaces are electropolished stainless steel to minimize absorption effects at ppb levels. Seals are gas compatible elastomers. A heated flow path is provided.

Commands are entered from the front panel and displayed on a backlit 25 line by 80 character liquid crystal display. Calibration sequences are stored in the internal microprocessor for recall by the keypad, optional RS-232 communications or optional status interface.

Humidification and pressurization options are available.

A purge circuit allows for nitrogen purge of the gas wetted surfaces to minimize contamination.

The Series 2014 is available in either a standard 19" rack mount or bench top configuration.



PRODUCT FEATURES AND BENEFITS



SOFTWARE

The Series 2014 has six primary routines accessible through on screen "soft" keys.

SPECIFICATIONS

Performance (as a percent of setpoint)

Accuracy              From 10 to 100% of Full Scale Flow
   Concentration:         ± 1.0%
   Flow:                  ± 1.0%
Repeatability:             0.05%
*Mass flow controllers are calibrated using a NIST Traceable Primary Flow Standard,
using a reference temperature of 0°C (32°F) and a reference pressure of 760 mm Hg (29.92 in. Hg).

Warm up time: 30 minutes

Mechanical

Gas connections
  1/4" Swagelok (or compatible fitting)

Operating Pressures
  Minimum:       10 psig (0.67 Bar)
  Recommended:   25 psig (1.68 Bar)
  Maximum:       75 psig (5.04 Bar)

Wetted Surfaces
  Tubing:       Electropolished 316 Stainless Steel
  MFC's:        Stainless Steel
  Seals:        Viton® (Optional - Kalrez®, Buna-N, Neoprene, Metal)

Operating temperature:    32° - 122° F   (0° - 50° C)
Performance temperature:  59°  - 95° F  (15° - 35° C)

Weight
  Minimum:      35 lbs. (16 Kg.)
  Maximum:      70 lbs. (32 Kg.)

Dimensions (w x h x d)
  Portable:  17" x 7" x 23.5"
             (43.18 cm x 17.78 cm x 59.69 cm)
  Rack:      19" x 7" x 23.5"
             (48.26 cm x 17.78 cm x 59.69 cm)

Electrical

  Standard:  115 VAC (100 to 130 VAC), 50/60 Hz
  Optional:  220 VAC (200 to 260 VAC), 50/60 Hz
  Current:     3 Amps (maximum)

Electronics

  Inmos T 400 series, 32 bit processor
  12 bit A/D and D/A conversion

Operating Modes

  Front panel membrane keypad
  Internal timer control
  RS-232 terminal mode / Remote computer control (optional)
  Status board interface (optional)
  Other communications options available

Data Output

  Optional RS-232 serial port / Parallel printer port interface


OPTIONS

Viton® and Kalrez® are registered trademarks of Dupont Dow Elastomers

About Environics | Products | Downloads | Tech Docs | Support
Contact Us | Customer Survey | Sales Reps | Distributors | Links | Home

Copyright © 2003 Environics, Inc. All Rights Reserved.
Environics is a registered trademark of Environics, Inc.
All other names mentioned are the property of their respective owners.