Series 6103
Ozone Transfer Standard / Multi-Gas Calibrator


DESCRIPTION

The Environics® Series 6103 Ozone Transfer Standard / Multi-Gas Calibrator is a computer controlled, state-of-the-art instrument for dynamic calibration of ambient air analyzers. The Series 6103 automatically performs zero, precision, span and multi-point calibrations using NO, NO2, SO2, CO, 03, hydrocarbons and other gases of interest. The 6103 meets or exceeds all U.S. Environmental Protection Agency requirements.

The Series 6103 consists of a single chassis supporting 2 thermal mass flow controllers, an ozone generation module, photometer, glass output manifold, mixing chamber, a reaction chamber for gas phase titration, and control electronics.

Series 6103 Picture Commands are entered from the front panel and displayed on a backlit 4 line by 20 character liquid crystal display. The instrument may also be remotely operated using contact closures or the RS-232 serial data interface, both are standard in the Series 6103.

The mass flow controllers are calibrated to a NIST (National Institute of Standards and Technology) traceable primary standard. The calibration data consists of a comparison of desired versus actual flow over the full dynamic range of the instrument with linear interpolation between points. Calibration data is stored in non-volatile memory and may be updated by the user with a suitable standard.

The internal ultra-violet (UV) based ozone generator is temperature controlled and includes a precision photo-optical feedback circuit to compensate for lamp aging effects providing stable ozone generation. The ozone generator is factory calibrated using a NIST traceable photometer standard.


PRODUCT FEATURES AND BENEFITS

User-friendly interactive software with plain language prompts is simple to use, reducing technician training time and virtually eliminating error.

Automatic calculation of dilution and span gas flows based on commanded concentration eliminates the need for manual computation and allows rapid transition from point to point.

Internally-stored mass flow controller calibration data improves accuracy by as much as a factor of ten and simplifies field recalibration. Mass flow controllers are factory calibrated at 11 points.

Internally-stored ozone generator calibration data insures linear, repeatable ozone generation without photometer control. Ozone generator performance meets U.S. EPA criteria for ozone transfer standards. The ozone generator is factory calibrated at 3 points (up to 11 points are available).

Ozone generator pressure compensation ensures repeatable and stable ozone generation at pressures other than the original calibration pressure. This is only important when the system is being used without the photometer closed-loop control.

The internal photometer measures the actual amount of ozone generated, and corrects for errors, using closed-loop PID control.

The photometer allows an external source of ozone to be analyzed and displayed on the screen. The external ozone is connector to a separate sample port, allowing for simultaneous gas blending while monitoring the external ozone source.

An automated leak test function is available to determine if the system has any internal leaks. Leaks are detected based on pressure decay.


SOFTWARE


SPECIFICATIONS

Performance*

  Accuracy
     Concentration:         ± 1.0% setpoint
     Flow:                  ± 1.0% setpoint
  Flow Repeatability:       ± 0.05% F.S.O.

  Warm up time:  30 minutes
* Performance specifications are valid when all Mass Flow Controllers are operating between 10% and 100% of full scale flow.
Mass flow controllers are calibrated using a NIST traceable Primary Flow Standard, using a Reference Temperature of 25°C (77°F) and a Reference Pressure of 760 mm Hg (29.92 in. Hg)
Ozone Generator
  Concentration Range:  0.02 - 0.5 ppm at 5 - 10 slpm
  Optional Ranges:      0.05 - 1.0 ppm at 5 - 10 slpm

UV Photometer
  Repeatability:   ±1 ppb
  Linearity:       0.3% F.S.O.
  Noise:           ±1 ppb
  Precision:        1 ppb

Mechanical

Inlets
  Balance: External 1/4" Swagelok*
  Span(s): External 1/4" Swagelok*

Outlet
    One external 1/4" Swagelok*

  *(or compatible fitting)

Operating Pressures at inlets
    Minimum:       20 psig (1.38 Bar)
    Nominal:       25 psig (1.72 Bar)
    Maximum:       30 psig (2.07 Bar)

Wetted Surfaces
    Tubing:          Teflon®
    MFCs:            Stainless Steel
    Seals:           Viton®
    Glass Chambers:  Pyrex®

Operating temperatures
    32° - 122° F (0° - 50° C)

Performance Temperature Range
    59° - 95° F (15° - 35° C)

Weight:  27 lbs.

Dimensions (w x h x d)
  Portable:  17" x 7" x 15"
  Rack:      19" x 7" x 15"
  (Optional:   19" x 7" x 24"  or  17" x 7" x 24")

Electrical

    Voltage: 100 to 250 VAC, 50/60 Hz
    Current:  2 Amps (maximum)

Operating Modes

  Front panel keypad
  RS-232 serial data interface
  I/O control (8 inputs / 8 outputs)

Data I/O

  RS-232 serial data interface
  I/O Control (8 inputs / 8 outputs)

OPTIONS