Product Data
Series 9100
Computerized Ambient Monitoring Calibration System
The Series 9100 consists of a single chassis supporting 2 thermal mass flow controllers, an ozone generation module, a mixing zone, a reaction chamber for gas phase titration, and control electronics.
Commands are entered from the front panel and displayed on a backlit 25 line by 80 character liquid crystal display. The instrument may also be operated using contact closures or an RS-232 serial data interface.
The mass flow controllers are calibrated to a NIST (National Institute of Standards and Technology) traceable primary flow standard. The calibration data consists of a comparison of desired versus actual flow over the full dynamic range of the instrument, with linear interpolation between points. Calibration data is stored in the non-volatile memory and may be updated by the user with a suitable standard.
The Series 9100 ozone generator is factory calibrated using a NIST traceable ozone standard. This temperature controlled, ultra-violet (UV) based ozone generator includes a precision photo-optical feedback circuit to compensate for lamp aging effects as well as built-in pressure compensation.
The Series 9100 is available in either a standard 19" rack mount or bench top configuration.
Accuracy From 10 to 100% of Full Scale Flow Concentration: ± 1.0% Flow: ± 1.0% Repeatability: 0.05% Warm up time: 30 minutes Response time: 30 seconds to ±2.0% of Setpoint*Mass flow controllers are calibrated using a NIST Traceable Primary Flow Standard,
Concentration Range: 0.5 - 1.25ppm at 10 slpm Accuracy: ±2% or ±0.005ppm at 10 slpm 30 day repeatability: ±2% or ±0.005ppm
Inlets
Balance: One external ¼" Swagelok*
Span(s): One external ¼" Swagelok*
Outlet
One external ¼" Swagelok*
* (or compatible fitting)
Operating Pressures (at inlets)
Minimum: 10 psig (0.67 Bar)
Recommended: 25 psig (1.68 Bar)
Maximum: 150 psig (9.87 Bar)
Wetted Surfaces
Tubing: Teflon
MFC's: Stainless Steel
Seals: Viton®
Operating temperature: 32° - 122° F (0° - 50° C)
Performance temperature: 59° - 95° F (15° - 35° C)
Weight
Minimum: 35 lbs. (16 Kg.)
Maximum: 40 lbs. (18 Kg.)
Dimensions (w x h x d)
Portable: 17" x 7" x 23.5"
(43.18 cm x 17.78 cm x 59.69 cm)
Rack: 19" x 7" x 23.5"
(48.26 cm x 17.78 cm x 59.69 cm)
Standard: 115 VAC (100 to 130 VAC), 50/60 Hz Optional: 220 VAC (200 to 260 VAC), 50/60 Hz Current: 3 Amps (maximum)
Inmos T 400 series, 32 bit processor 12 bit A/D and D/A conversion
Front panel membrane keypad Internal timer control Optional RS-232 serial data interface Optional Status board interface Other communications options available
Optional RS-232 port serial data interface Optional Parallel printer port (Centronics compatible)
|
About Environics | Products | Downloads | Tech Docs | Support Contact Us | Customer Survey | Sales Reps | Distributors | Links | Home Environics is a registered trademark of Environics, Inc. All other names mentioned are the property of their respective owners. |