Environics

Series 9100
Computerized Ambient Monitoring Calibration System


DESCRIPTION

The Environics® Series 9100 Computerized Ambient Monitoring Calibration System is an advanced microprocessor controlled instrument for dynamic calibration of ambient air analyzers. The Series 9100 automatically performs zero, precision, span and multi-point calibrations using NO, NO2, SO2, CO, O3, hydrocarbons and other gases of interest. The Series 9100 exceeds all U.S. Environmental Protection Agency requirements.

The Series 9100 consists of a single chassis supporting 2 thermal mass flow controllers, an ozone generation module, a mixing zone, a reaction chamber for gas phase titration, and control electronics.

Commands are entered from the front panel and displayed on a backlit 25 line by 80 character liquid crystal display. The instrument may also be operated using contact closures or an RS-232 serial data interface.

The mass flow controllers are calibrated to a NIST (National Institute of Standards and Technology) traceable primary flow standard. The calibration data consists of a comparison of desired versus actual flow over the full dynamic range of the instrument, with linear interpolation between points. Calibration data is stored in the non-volatile memory and may be updated by the user with a suitable standard.

The Series 9100 ozone generator is factory calibrated using a NIST traceable ozone standard. This temperature controlled, ultra-violet (UV) based ozone generator includes a precision photo-optical feedback circuit to compensate for lamp aging effects as well as built-in pressure compensation.

The Series 9100 is available in either a standard 19" rack mount or bench top configuration.


PRODUCT FEATURES AND BENEFITS

  • 25 line by 80 character display permits editing and viewing of an entire seven-day schedule of calibration events in worksheet format.
  • Automatic calculation of dilution and span gas flows based on commanded concentration eliminates the need for manual computation and allows rapid transition from point to point.
  • Internally-stored mass flow controller calibration data improves accuracy by as much as a factor of ten and simplifies field recalibration.
  • Internally-stored ozone generator calibration data insures linear, repeatable ozone generation without photometer control. Ozone generator performance exceeds U.S. EPA criteria for ozone transfer standards.
  • Ozone generator pressure compensation insures repeatable ozone generation at a flow rate and back pressure other than that at which it was originally calibrated.

SOFTWARE

The Series 9100 has six primary routines.

  • Concentration Mode: User enters Target Output Gas Concentration (ppm or %) for the span gas. The actual concentration is displayed during mixing.
  • Flow Mode: User enters Target Flow Rate (cc's or liters per minute) for the span gas and dilution gas. Actual flow rates are displayed after mixing is initiated.
  • Volts Mode: User enters a command voltage to the mass flow controller and the ozone generator. The actual voltage is displayed during operation.
  • Program Mode: Permits multi-event programming and unattended automatic operation of the instrument over a seven-day schedule.
  • Calibrate Mode: User enters new calibration data for each mass flow controller and ozone generator. (Can be password protected)
  • Maintain Ports: User enters the name of the span gas in the source cylinder, its concentration (ppm) and the port to which the cylinder is connected.

SPECIFICATIONS

  • Mass Flow Controllers (as a percent of setpoint)*

    Accuracy
    From 10 to 100% of Full Scale Flow
    Concentration: ± 1.0%
    Flow: ± 1.0%
    Repeatability: 0.05%

    Warm up time: 30 minutes
    Response time: 30 seconds to ±2.0% of Setpoint

    *Mass flow controllers are calibrated using a NIST Traceable Primary Flow Standard,
    using a reference temperature of 25°C (77°F) and a reference pressure of 760 mm Hg (29.92 in. Hg).

    Ozone Generator:

    Concentration Range: 0.5 - 1.25ppm at 10 slpm
    Accuracy: ±2% or ±0.005ppm at 10 slpm
    30 day repeatability: ±2% or ±0.005ppm

    Mechanical

    Inlets
    Balance: One external ¼" Swagelok*
    Span(s): One external ¼" Swagelok*

    Outlet
    One external ¼" Swagelok*
    * (or compatible fitting)

    Operating Pressures (at inlets)
    Minimum: 10 psig (0.67 Bar)
    Recommended: 25 psig (1.68 Bar)
    Maximum: 150 psig (9.87 Bar)

    Wetted Surfaces
    Tubing: Teflon
    MFC's: Stainless Steel
    Seals: Viton®

    Operating temperature: 32° - 122° F (0° - 50° C)
    Performance temperature: 59° - 95° F (15° - 35° C)

    Weight
    Minimum: 35 lbs. (16 Kg.)
    Maximum: 40 lbs. (18 Kg.)

    Dimensions (w x h x d)
    Portable: 17" x 7" x 23.5"
    (43.18 cm x 17.78 cm x 59.69 cm)

    Rack: 19" x 7" x 23.5"
    (48.26 cm x 17.78 cm x 59.69 cm)

    Electrical

    Standard: 115 VAC (100 to 130 VAC), 50/60 Hz
    Optional: 220 VAC (200 to 260 VAC), 50/60 Hz
    Current: 3 Amps (maximum)

    Electronics

    Inmos T 400 series, 32 bit processor
    12 bit A/D and D/A conversion

    Operating Modes

    Front panel membrane keypad
    Internal timer control
    Optional RS-232 serial data interface
    Optional Status board interface
    Other communications options available

    Data Output

    Optional RS-232 port serial data interface
    Optional Parallel printer port (Centronics compatible)

    OPTIONS

    • Rack Mount
    • Extra Gas Inlets (Limit-six total)
    • MFC
    • RS-232 Serial Data Interface
    • Solenoid Valve on Output
    • Status Board
    • Permeation Oven
    Viton® is a registered trademark of Dupont Dow Elastomers
© Environics, Inc. All rights reserved. Phone: (860) 872-1111